Study of the dosimetric properties of CaSO4:Dy using OSL technique

P.B.R. Gasparian1, A.L.M.C. Malthez2, E.M. Yoshimura3, N.K. Umisedo3 and L.L. Campos1 Published 29 April 2022 • © 2022 IOP Publishing Ltd and Sissa Medialab

Click to access the paper

Abstract

Calcium sulfate doped with dysprosium (CaSO4:Dy) has been used for personal dosimetry using the thermoluminescence (TL) technique for decades and its TL properties are well-known. Considering the advantages of the OSL technique (relatively simple and faster readout and detector reusability) and the potential application of CaSO4:Dy as an OSL detector, this study aimed to investigate the dosimetric properties of CaSO4:Dy using the OSL technique. The present study was carried out using CaSO4:Dy pellets prepared at IPEN irradiated and read in a TL/OSL Risø reader with blue light stimulation and Hoya U-340 filter. OSL dosimetric properties essentially dose response, reproducibility, and fading characteristics were evaluated. Moreover, TL and OSL signals were compared to study the correlation between OSL signals and TL peaks. CaSO4:Dy OSL response is linear from 0.1 to 10 Gy and shows reproducibility better than 5%. The low-temperature TL peak is unstable; however, the OSL signal seems to be associated with higher-temperature peaks.